Test Methods
Electrical and Electronic Test Equipment
Method Used: CATS Internal Test Methods based on Euramet-CG15 and Fluke Calibration:Philosophy in Practice Second Edition
Accuracy Spec: Manufacturer's specification, unless requested otherwise by the user
Decision Rule: The sum of the absolute value of the correction and magnitude of uncertainty must not be greater than accuracy specification at any point
Oscilloscopes
Method Used: CATS Internal Test Methods based on Euramet-CG7
Accuracy Spec: Manufacturer's specification, unless requested otherwise by the user
Decision Rule: The sum of the absolute value of the correction and magnitude of uncertainty must not be greater than accuracy specification at any point
Infrared Thermometers and Thermal Imagers
Method Used: CATS Internal Test Methods based on MSL Technical Guide 22, MSL Technical Guide 26, MSL Technical Guide 29
Accuracy Spec: Manufacturer's specification, unless requested otherwise by the user
Decision Rule: The sum of the absolute value of the correction and magnitude of uncertainty must not be greater than accuracy specification at any point
Digital Pressure Instruments
Method Used: MSA Test Method 1
Accuracy Spec: Manufacturer's specification, unless requested otherwise by the user
Decision Rule: The sum of the absolute value of the correction and magnitude of uncertainty must not be greater than accuracy specification at any point
MECHANICAL Pressure Instruments
Method Used: MSA Test Method 2
Accuracy Spec: Manufacturer's specification, or based on dial size as defined in MSA Test Method 2
Decision Rule: No correction or uncertainty can be greater than the accuracy specification at any point, and reference uncertainty must be within limits as defined in MSA Test Method 2.